TL-980 Standard Test Lead Kit for 98x Insulation Resistance (IR) Tester
Expanding the possibilities of High Voltage Resistance Measurement
The new 981i and 983i Teraohmmeters are designed to tackle the toughest High Resistance measurement applications. Demanding applications that most other IR testers won’t measure up to. So when it’s time to test higher voltage IR values like today’s electric vehicle systems and higher voltage solar arrays — look to Vitrek to deliver the IR tester you need for your production line.
What sets the Vitrek 98x Series IR testers apart from the others?
For starters, these units are based on Vitrek’s proven DSP technology — so they have the ability to work into capacitive loads where most others do not. And when it comes to output voltage range, Vitrek delivers — with up to 6.5 KVDC for the 981i and up to 11KV with the 983i.
Need to Perform an Insulation Resistance Test on a Multi-Conductor Cable Harness?
The 981i and 983i have the ability to directly control up to four 64 channel HV scanners, right out of the box. That is up to 256 test points and using a PC with Vitrek’s Q-Test software you can expand the count up to 1020 test points. The HV Switching System of choice is the Vitrek 964i, which can hold eight 8-channel switching cards — available in 7, 10 and 15KV ratings.
In a never ending race for higher efficiency, electric vehicles and solar panel arrays are now operating at 2.5KV.
This is no place for a 1KV tester, it’s time to step up to a 98x!
Product Features and Benefits
Specifications
General Specifications
Specifications: Valid after 15 minute warm-up, for 1 year from last external calibration, and for ambient temperature within +/- 2 Degrees C of last performed ZERO operation.
Ethernet: High speed, high noise immunity LAN interface
RS232 Interface: Selectable baud: 9600, 19200, 38400, 57600 or 115200, full handshake
VICL Interfaces: Provided for control of HV Scanners
Digital I/O Interface: Provides 4 digital inputs and 5 digital outputs. Functions include, Start/Stop, Testing, Pass/Fail, HV Present, Safety Interlock, Dwell Timer
Optional GPIB: Option GPIB-9 adds GPIB interface capability
Operating Temperature: 0°C to 50°C
Humidity: <85% RH (non-condensing)
Power: 105-265Vrms, 45-65Hz, 500VA Max
Dimensions: 89mm H x 432mm W x 457mm D (3.5″ H x 17″ W x 18″ D)
Weight: 5Kg (11 lb.) Net / 7Kg (15 lb.)
Accessories: Alligator test leads, NIST traceable calibration certificate with no data, power cord and operator’s manual. Limited ISO 17025 cal cert with data & uncertainties available.
DC Output Voltage
981i: 30V to 6500V
Accuracy: +/- (0.25% rdg + 0.1V)
981i with option HSS2: 30V to 5000V
Accuracy: +/- (0.25% rdg + 0.1V+1V/mA)
983i: 60V to 11000V
Accuracy: +/- (0.25% rdg + 0.2V)
Output Resolution: 1V
Loading: up to 4mA charging, 2mA continuous
Current Measurement
981i & 983i RETURN current only, 1pA to 4mA
Accuracy (Resistive Load up to 1000pF): +/- (0.5%rdg + 10pA + 2pA per kV))
Accuracy (Capacitive Load 1000pF to 50nF): +/- (0.5%rdg + 100pA + 20pA per kV))
Accuracy (Capacitive Load >50nF): +/- (0.5%rdg + 1nA + 200pA per kV))
Option HSS2 HV Output current, 1nA to 4mA (for grounded loads)
Accuracy (Resistive Load up to 1000pF), C<1000pF): +/- (0.5%rdg + 5nA + 1nA per kV)) Accuracy (Capacitive Load 1000pF to 50nF): +/- (0.5%rdg + 7.5nA + 2.5nA per kV)) Accuracy (Capacitive Load >50nF): +/- (0.5%rdg + 10nA + 5nA per kV))
Resistance Measurement
NOTE: the use of the OFFSET (or ZERO) capability in either AUTO or MANUAL TEST modes is not required to meet these specifications – the OFFSET capability is intended to allow the user to offset externally produced leakages.
Transfer Measurement
Comparative Measurement Capability (measurement of multiple values within 10% at the same test voltage, in the same environmental conditions within +/-1C, within a 1 hour period, with no intervening ZERO operations, and constant power)
Uncertainty (Resistive Load): +/- (0.05%rdg + (R/(5Tohm per kV))%rdg + (1/V)%rdg + 2ohms per kV) for R < 150TohmTiming
Charge Time: 1 second minimum charge time, 9999 sec max.
Dwell Time: 0.1s to 9999 days, 0.1 sec resolution
Ordering & Accessories
Below prices in US$ are for destinations in North America with terms of prepay, net 30 days (OAC) and accepted credit cards. For shipment beyond North America or terms beyond 30 days, please contact the factory for a quotation. If you require rush/expedited shipment please contact Vitrek at (858) 689-2755 or email info@vitrek.com
98x IR Tester
Thumbnail | Name | Short description | Price | Buy |
---|---|---|---|---|
981i | 981i 6.5KV Teraohmmeter/IR Tester | $6,672.00 | ||
983i | 983i 11KVDC Â Teraohmmeter/IR Tester, includes TL-980 and NIST traceable Calibration certificate. | $7,785.00 |
98x IR Tester Accessories
Thumbnail | Name | Short description | Price | Buy |
---|---|---|---|---|
98X-230V | Factory Set 230V Line Operation. Order this no charge option when powering the unit from 210-250VAC. | $0.00 | ||
95X-230V | Factory Set 230V Line Operation. Order this no charge option when powering the unit from 210-250VAC. | $0.00 | ||
QTE-9 | QT Enterprise Sofware License Key for use with 98x an 95x products. | $1,798.00 | ||
TL-980 | TL-980 Standard Test Lead Kit for 98x Insulation Resistance (IR) Tester | $317.00 | ||
HSS-2 | DUT Isolation Option. Option HSS-2 adds the ability to measure DC breakdown and/or leakage into a grounded DUT with higher resolution than HSS-1. Maximum current capabilities are further reduced. (This option may be fitted in a 951i, 952i, or 956i Hipot Tester or 981i Teraohmmeter/IR Tester) | $948.00 | ||
RFS-95 | Remote Start Foot Switch | $247.00 | ||
GPIB-9 | Optional GPIB Interface | $626.00 | ||
RM-1 | Rack Mount Kit | $95.00 |
Rentals
The 98x Series Teraohmmeter/IR Tester and accessories are available for rent. Please contact the factory for a rental quotation on the items and period of time you desire. Vitrek offers a generous rental allowance for rentals that are converted to a purchase and rentals of 12 months come with a 10% buyout.